MODEL |
JTH-20L-70-
CPA/W |
JTH-30L-70-
CPA/W |
JTH-64L-70-
CPA/W |
JTH-80L-70-
CPA/W |
JTH-150L-70-
CPA/W |
Chamber Design |
Internal
Volume |
L |
20 |
30 |
64 |
80 |
150 |
Internal
Dimensions (cm) |
W |
34 |
30 |
40 |
40 |
50 |
H |
15 |
40 |
40 |
50 |
60 |
D |
40 |
25 |
40 |
40 |
50 |
External
Dimensions (cm) |
W |
50 |
85 |
56 |
60 |
70 |
H |
86 |
100 |
105 |
146 |
156 |
D |
86 |
65 |
85 |
123 |
133 |
Temperature Test Parameters |
Temperature
Range |
°C |
-70 ℃/-40 ℃~85 ℃/120 ℃/150 ℃/180 ℃ or customized |
Heating Rate |
°C/min |
1 ℃/min (empty
load) or customized |
Cooling Rate |
°C/min |
3 ℃/min (empty load) or customized |
Temperature
Fluctuation |
°C |
±0.5 |
±0.5 |
±0.5 |
±0.5 |
±0.5 |
Temperature
Uniformity |
°C |
≤ 2.0 (unloaded, constant state) |
Temperature Deviation |
°C |
±2 |
±2 |
±2 |
±2 |
±2 |
Humidity Test Parameters |
Humidity Range |
%RH |
20%~95% |
Humidity Deviation |
%RH |
± 3.0% RH when>75% RH, ± 5.0% RH when ≤ 75% RH |
Humidity Fluctuation |
%RH |
±2 |
±2 |
±2 |
±2 |
±2 |
Humidity Uniformity |
%RH |
5 |
5 |
5 |
5 |
5 |
Power supply and Connection (Subject To Actual Usage Configuration) |
Configure Interface |
Standard RS485 port/USB, optional WLAN and APP functions, with power on test hole |
Rated Voltage |
V/Hz |
AC380V 50HZ |
PLC Function |
Automatically call group PID parameters |
Protector |
Imported over temperature protection, dry burning protection, humidification protection, overload protection, high voltage protection, leakage protection |
Control Mode |
PID+SSR |
Environmental Temperature/Humidity For Use |
5℃~35℃;≤ 85%RH |
System Components |
Material Science |
Shell Material |
Cold rolled steel plate electrostatic double-sided spray coating |
Inner Liner Material |
SUS # 304 stainless steel plate |
Thermal Insulation Material |
150mm hard polyurethane foam insulation layer |
Refrigeration System |
Cooling Method |
Mechanical binary freezing method (air-cooled/water-cooled) |
Compressor |
Full density vortex compressor |
Refrigerant |
R404a, R23 environmentally friendly refrigerant |
Heating System |
Finned nickel chromium alloy electric heating tube |
Circulating Fan |
Adopting a multi blade centrifugal circulating fan with reinforced shaft and aluminum alloy rotating blades that are resistant to high and low temperatures |
Door Observation Window |
A 3-layer vacuum glass observation window for heat and sweat prevention;
Door frame electric heating (automatic adjustment) anti frost and anti condensation device |
Temperature Sensor |
Platinum resistance PT100 |
Humidity Sensor |
Dry wet bulb sensor |
Temperature/Humidity Controller |
High precision resistive touch screen, can be controlled by a USB external mouse;
Historical curve record, PID+SSR/SCR automatic bidirectional synchronous output |
Noise |
dB(A) |
≤ 70dB (A-level sound level) |
Net Weight |
KG |
Based on the actual customized weight, the ground load requirement for the venue is ≥ 1000kg/㎡ |
Enclosure |
Double opening imported hinged door, 1 observation window, 1 built-in slope, customized operation hole, overflow hole, lighting, air pressure balance device, etc |